Swoop® Portable MR Imaging® System

Sequences

Swoop Sequences

We invite you to learn more about Swoop system T1, T2, FLAIR, and DWI sequences.

T1 (Standard) / T1 (Gray/White)

Resolution: 1.6 × 1.6 × 5 mm

Acquisition type; orientations: 3D; axial, sagittal, coronal

Kspace trajectory: Cartesian, weighted coverage

Image reconstruction: Kalman filter noise suppression, line noise suppression, advanced reconstruction, phase-sensitive parallel imaging reconstruction, navigator

Post-processing: Geometric distortion correction, intensity correction

Additional info (T1 Gray/White): Similar to T1-weighted gradient-echo-based imaging sequences on higher field MRIs. Optimized for visualizing extra-axial pathology.

Additional info (T1 Standard): Similar to T1-weighted spin-echo-based imaging sequences on higher field MRIs. Optimized for visualizing intraparenchymal pathology.

T1 (Standard)
T1 (Gray/White)

T2 / Fast T2

Resolution: T2: 1.5 × 1.5 × 5 mm; Fast T2: 1.6 × 1.6 × 5 mm

Acquisition type; orientations: 3D; axial, sagittal, coronal

Kspace trajectory: Cartesian, weighted coverage

Image reconstruction: Kalman filter noise suppression, line noise suppression, advanced reconstruction, phase-sensitive parallel imaging reconstruction, navigator

Post-processing: Geometric distortion correction, intensity correction

Additional info: Fast T2 shortens the scan time by approximately 50%.

T2
Fast T2

Fluid-Attenuated Inversion Recovery (FLAIR)

Resolution: 1.6 × 1.6 × 5 mm

Acquisition type; orientations: 3D; axial, sagittal, coronal

Kspace trajectory: Cartesian, weighted coverage

Image reconstruction: Kalman filter noise suppression, line noise suppression, advanced reconstruction, phase-sensitive parallel imaging reconstruction, navigator

Post-processing: Geometric distortion correction, intensity correction

FLAIR

Diffusion-Weighted Imaging (DWI) and Apparent Diffusion Coefficient (ADC) Maps

Resolution: 2.4 × 2.4 × 6 mm

Acquisition type; orientations: 3D; axial

Associated options: ADC mapping, b-value = 900 s/mm2

Kspace trajectory: Cartesian, weighted coverage

Image reconstruction: ADC map output, Kalman filter noise suppression, line noise suppression, FISTA reconstruction, navigator

Post-processing: Geometric distortion correction, intensity correction

DWI
ADC

Learn More

click to read about The Clinical and Business Value of Portable Point‑of‑Care MRI - opens in new window
{disply_title}
White Paper

The Clinical and Business Value of Portable Point‑of‑Care MRI

click to read about Clinical Use Cases for the Hyperfine Swoop® Portable MR Imaging System - opens in new window
{disply_title}
White Paper

Clinical Use Cases for the Hyperfine Swoop® Portable MR Imaging System

click to read about Details and Specs - opens in new window
{disply_title}
Link

Details and Specs

Review Swoop® system details and specifications.

click to read about FAQs - opens in new window
{disply_title}
Link

FAQs

Read the most frequently asked questions about the Swoop® system.

click to read about Where to Find Swoop® - opens in new window
{disply_title}
Link

Where to Find Swoop®

Point-of-care neuroimaging near you.

This website stores cookies on your computer. These cookies are used to improve your website experience and provide more personalized services to you, both on this website and through other media. To find out more about the cookies we use, see our Privacy Policy.

If you click 'Decline', we won't track your information when you visit our site. But in order to comply with your preferences, we'll have to use just one tiny cookie so that you're not asked to make this choice again.