We invite you to learn more about Swoop system T1, T2, FLAIR, and DWI sequences.
T1 (Standard) / T1 (Gray/White)
Resolution: 1.6 × 1.6 × 5 mm
Acquisition type; orientation(s): 3D; axial, sagittal, coronal
K-space trajectory: Cartesian, weighted coverage
Image reconstruction: Interference noise suppression, advanced gridding
Post-processing: Geometric distortion correction, intensity correction, advanced denoising
Additional info (T1 Gray/White): Similar to T1-weighted gradient-echo-based imaging sequences on higher field MRIs.
Additional info (T1 Standard): Similar to T1-weighted spin-echo-based imaging sequences on higher field MRIs.
T2 / Fast T2
Resolution: T2: 1.5 × 1.5 × 5 mm; Fast T2: 1.6 × 1.6 × 5 mm
Acquisition type; orientation(s): 3D; axial, sagittal, coronal
K-space trajectory: Cartesian, weighted coverage
Image reconstruction: Interference noise suppression, advanced gridding
Post-processing: Geometric distortion correction, intensity correction, advanced denoising
Additional info: Fast T2 is approximately 60% shorter compared to T2.
Fluid-Attenuated Inversion Recovery (FLAIR)
Resolution: 1.7 × 1.7 × 5 mm
Acquisition type; orientation(s): 3D; axial, sagittal, coronal
K-space trajectory: Cartesian, weighted coverage
Image reconstruction: Interference noise suppression, advanced gridding
Post-processing: Geometric distortion correction, intensity correction, advanced denoising
Diffusion-Weighted Imaging (DWI) and Apparent Diffusion Coefficient (ADC) Maps
Resolution: 2.4 × 2.4 × 6 mm
Acquisition type; orientation(s): 3D; axial
Associated options: ADC mapping, b-value = 900 s/mm2
K-space trajectory: Cartesian, weighted coverage
Image reconstruction: Interference noise suppression, advanced gridding, motion correction
Post-processing: Geometric distortion correction, intensity correction, advanced denoising
Please click here for region-specific clearances.